Titan Instructions for Use - EN Page 98
Reflex Calibration Standards and Spectral Properties:
Specifications for stimulus and audiometer signals are made to follow IEC 60645-5
ISO 389-1 for TDH39 and ISO 389-2 for CIR 55.
As “Broad band noise” specified in IEC 60645-5, but
with 500 Hz as lower cut-off frequency.
Uniform from 500 Hz to 1600 Hz, ±5 dB re. 1000 Hz
Uniform from 1600 Hz to 10KHz, ±5 dB re. 1000 Hz
As “Broad band noise” specified in IEC 60645-5, but
with 500 Hz as lower cut-off frequency.
Uniform from 500 Hz to 1600 Hz, ±10 dB re. 1000 Hz
Uniform from 1600 Hz to 4000 Hz, ±10 dB re. 1000
The actual sound pressure level at the eardrum will
depend on the volume of the ear. See Table 2 for
The risk of artifacts at higher stimulus levels in reflex measurements are minor and will not activate the
reflex detection system