Titan Instruction for Use EN Page 88
Cleaning the Probe Tip
In order to secure correct measurements it is important to make sure that the probe system is kept clean at
all times. Therefore please follow the below illustrated instructions on how to remove e.g. cerumen from the
small acoustic and air pressure channels of probe tip.
Probe, Short Extension
Cable, Preamplifier
Clinical Extension Cable
1. Unscrew the probe cap.
2. Remove the probe tip.
3. Thread the stiff end of the cleaning
floss into one of the tubes.
4. Pull the cleaning floss completely
through the probe tip tube. Clean each
of the tubes as required. Discard floss
after use.
5. Assemble the probe again.
Notice:
Do not use the cleaning tool to clean the probe base. This will destroy the filters.