Titan Instructions for Use - EN Page 102
<5% UT
(>95% dip in UT) for 5 sec
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
residential environment.
Note: UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration — electromagnetic immunity
The TITAN is intended for use in the electromagnetic environment specified below. The customer or the user of the TITAN should assure
that it is used in such an environment,
IEC / EN 60601
test level
Electromagnetic environment – guidance
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
3 Vrms
150kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Portable and mobile RF communications
equipment should be used no closer to
any parts of the TITAN , including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
800 MHz to 2,5
GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, (a) should
be less than the compliance level in
each frequency range (b)
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
(a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
TITAN is used exceeds the applicable RF compliance level above, the TITAN should be observed to verify normal operation, If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the TITAN .
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.