Titan Instructions for Use - EN Page 83
Screening results can be PASS, REFER or INCOMPLETE and
are indicated above the measurement as soon as the result is
available. If the “Enabled Pass/Refer” checkbox is not checked for
the selected protocol, no labeling will appear.
The statistical significance of the detected result is dependent upon
a combination of the following user defined settings in the protocol
setup: Test time, Stimulus level. SNR, Recording Window, Min
Total OAE, Min Reproducibility, Min TE level, Number of bands
required for a pass, Mandatory bands required for a pass.