SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1143482 13 Perfecto
2
Series
Change in maximum
recommended flow when
back pressure of 7kPa is
applied:
0.7 L/min
Filters: Cabinet, Outlet HEPA and Compressor Inlet
Safety System: Current overload or line surge shutdown. High
temperature compressor shutdown. High
Pressure Alarm w/compressor shutdown. Low
Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO
2
Oxygen
System (SensO
2
Model) Low Flow Alarm.
Width: 15 in ± 3/8 in (38.1 cm ± 1 cm)
Height: 23 in ± 3/8 in (58.4 cm ± 1 cm)
Depth: 12 in ± 3/8 in (30.5 cm ± 1 cm)
Weight: 43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto
2
Models
45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto
2
W Models
40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto
2
V Model
Shipping Weight: 48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto
2
Models
50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto
2
W Models
45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto
2
V Model
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity Perfecto
2
Models and Perfecto
2
V Model
50°F - 90°F (10°C - 32°C) at 20-60% relative
humidity Perfecto
2
W Models
Exhaust: Less than Ambient + 35°F (+19°C)
Oxygen Output
Temperature:
Less than Ambient + 6°F (+3°C)
Cabinet: Impact Resistant flame-retardant plastic cabinet
that conforms to UL94V-0.
Regulatory Listing: Double Insulated Product ETL certified
complying with UL1431 and UL1097 ETL certified
to CSA C22.2 No. 68.
Electrical: No extension cords.