SECTION 6—MAINTENANCE
Part No 1143482 25 Perfecto
2
Series
FIGURE 6.2 Preventive Maintenance Record
ON EACH INSPECTION
MODEL NO. I R C SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY 4,380 HOURS
Check Oxygen Concentration
(UNITS WITHOUT SensO
2
)
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
(
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE:
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Check Oxygen Concentration
UNITS WITH SensO
2
)
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST