Why does Invacare Perfecto 2 SeriesIRC5PO2AW Medical Equipment have low concentration even after checking purity?
- TThomas SmithSep 12, 2025
If your Invacare Medical Equipment has low concentration and you've already checked for oxygen purity, the flowmeter may be the issue. If the flowmeter is opened beyond the maximum flow rate, return it to the maximum setting. If there is a cracked or broken fitting, replace the fitting. If the input tubing is leaking or loose, repair or replace it. Additionally, the P.E. valve could be faulty, with a bad coil or a restrictor blockage, in which case you should replace the P.E. valve. Finally, inspect the P.C. board restrictor tubing for kinks or tears and replace the P.C. board if necessary.