Usage
Pos.
DescriptionFunction
D
PowerfailureindicatorRedLEDindicatorashesandaudiblebuzzersoundsifthesystemhas
nopowerconnected.Thiscouldbeintheeventofapowerfailure,for
examplecausedbypowercordunpluggedorpoweroffwhencontrol
unitisworking.Whenpowerisrestored,controlunitwillautomatically
startworking.
E
Pressurefailure/Lowpressure
indicator
RedLEDindicatorashesandaudiblebuzzersoundsifthesystem
recognisesalowpressurefailure.
Operation
1.Thepowerindicatorwillashandcontrolunitwillenterstaticmodeinitiallyuntilpressurereaches15mmHg.Thencontrol
unitwillenteralternatemodeandpowerindicatorwillremainon.Thecycletimeissetat10min,andthepressureis
setat60±3mmHg.
Therstinationofnewmattress:theindicatorfunction(visibleandaudibleindicators)willactivateifthemattressdoes
notinatecompletelywithin15minutes.
Operationfailures
•Duringcontrolunitoperation,lowpressureindicatorwillactivatewithin1minutesifairhosesaredisconnectedor
aircellsarebroken.
•Inthecaseofabnormalalternateornoalternate,theaudiblesignalwillbeactivatedandalternatefailureindicatorwillash.
5.3CPRprocedure
TheInvacare®SoftformActive®2Rxmattresshasbeen
fullytestedtocomplywiththecurrentCPRstandardof5-
6cmcompressiondepth.Thiswasachievedatallstages
ofination/deation.
1632862-A15