Invacare®SoftformPremierActive2withRXpump
6.2.2CleaningIntervals
IMPORTANT!
Regularcleaninganddisinfectionenhances
smoothoperation,increasestheservicelifeand
preventscontamination.
Cleananddisinfecttheproduct
–regularlywhileitisinuse,
–beforeandafteranyserviceprocedure,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser .
IMPORTANT!
–Keepacleaningrecordaspartofcleaningthe
system.
6.2.3CleaningInstructions
IMPORTANT!
–Theproductdoesnottoleratecleaningin
automaticwashingplants,withhigh-pressure
cleaningequipmentorsteam.
Acleaningrecordshouldbekeptaspartofcleaning
thesystem.
Cleaningcontrolunit
1.Wipedownthecontrolunitcasingandhosettings
withadampclothandsuitabledetergent.
2.Usinganylonbrush,gentlycleanallcrevicesasthey
canharbormicroorganisms.
3.Wipeoffthecontrolunitcasingandhosettingswitha
dampclothtoremovealldetergent.
4.Airdryalltreatedsurfaces.
Cleaninghangers
1.Wipedowntheexteriorofthehangersperiodically,
usingadampclothandsuitabledetergent.
Cleaningcovers
(Removalofcontaminantssuchasdustandorganicmatter)
1.Removeallcoversforlaundering.
2.Launderthecoverswithmaximumtemperatureas
statedonproductlabel,usingadiluteddetergent
solution(Instructionsonlabel).
IMPORTANT!
Washingathighertemperatureswillcause
shrinkage.
Dryingcovers
1.Hangcoversfromalineorbaranddripdryinaclean
indoorenvironmentortumbledryonalowheatsetting.
IMPORTANT!
–Tumbledrysettingmustnotexceed104°F.
–Donottumbledryforlongerthan10minutes.
–Drythoroughlybeforeuse.