Invacare®SoftformPremierActive2withRXpump
8Troubleshooting
8.1Identifyingandrepairingdefects
Thereareaudioandvisualalarmspresentonthecontrolunit.
Problem/Alarm
Cause
Solution
MattressCPRhosedisconnected.ConnectCPRhoseconnector,lockitinplace.
Powercableandfusehasbeenchecked,control
unitdoesnotoperate.
SendcontrolunitbacktoInvacareforrepair.
Mattressnotinating
(notalternating
properly).
Alternatingsystem
/Timingfailure
indication
Majorleakinaircellorcompleteairinsert.Replaceleakingaircell.
CPRhoseortubeconnectorskinkedorsplit.UnkinkorreplacesplitCPRhoseortube
connectors.
Notalternating,rotorfailure.SendcontrolunitbacktoInvacareforrepair.
Noair(controlunitfailure).
SendcontrolunitbacktoInvacareforrepair.
Controlunitoff.Checkpowersource,turnuniton.
Nopower/
Powerfailure
indication
Powercorddisconnected.Connectpowercordandensurethepowersource
ison.
Nopowerinthepoweroutlet.
Switchonthepoweroutlet(ifpoleswitch
available).
Havethepoweroutletrepairedbyanelectrician.