EasyManua.ls Logo

Invacare Softform Premier Active 2 - Technical Data; Mattress Specifications; Control Unit Specifications

Invacare Softform Premier Active 2
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Invacare®SoftformPremierActive2withRXpump
9TechnicalData
9.1MattressSpecications
Dimensions[inch]
Product
LengthWidthHeight
Aircellheight
[inch]
Maximum
userweight
[lbs]
Weightof
product[lbs]
1
8036500
SoftformActive2Rx
80,8439,42,48
63
650
35
9.2ControlUnitSpecications
SupplyVoltage
120V,60Hz
RatedInputCurrent
1A
FuseRating
1A
NoiseLevelbelow24dB
ClassicationClassIITypeBF
CycleTime
10min,A/B+/-1min
Dimensions(LengthxWidthxHeight)
10.8inx6.1inx4.1in
Weight3.9lbs
AirFlow8lpm
OperatingPressure
60mmHg(8kPa)
1.Weightcanvarydependingonmattresssize.Averageweightusedasanindication.
24
1632862-A