TechnicalData
Electromagneticimmunity
Immunitytest
IEC60601testlevelCompliancelevelElectromagneticenvironmentguidance
Electrostaticdischarge
(ESD)
IEC61000-4-2
±8kVcontact
±2kV,±4kV,±8kV,
±15kVair
±8kVcontact
±2kV,±4kV,±8kV,
±15kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electricalfast
transient/burst
IEC61000-4-4
±2kVforpowersupply
lines
±1kVforInput/output
lines
±2kVforpowersupply
lines
±1kVfor
interconnectingcable
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Surge
IEC61000-4-5
±1kVlinetoline
±2kVlinetoearth
±1kVlinetolineMainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Interruptionsand
voltagevariationson
powersupplyinput
lines
IEC61000-4-11
<5%U
T
(>95%dipin
U
T
)
for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipin
U
T
)
for5sec
<5%U
T
(>95%dipin
U
T
)
for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipin
U
T
)
for5sec
Mainspowerqualityshouldbethatofa
typicalcommercialorhospitalenvironment.
Iftheuserofthedevicerequirescontinued
operationduringpowermainsinterruptions,itis
recommendedthatthedevicebepoweredfrom
anuninterruptiblepowersupplyorabattery.
1632862-A