InvacarSoftform®PremierActive2S
7Maintenance
7.1Inspection(multipleuse)
Itisrecommendedthatthemattressischecked(foam,air
insertandcover)forstrike-through(thismayincludeuid
ingress,stains,ripsordamage)afterthereleaseofeach
patient,afterendingoftheperiodofuseoronaminimum
monthlybasis(dependingonwhichoccursrst)byasuitably
qualiedandcompetentperson.
Checkmattress(multipleuse)
1.Unzipthecovercompletely.
2.Checkforanystainingonthewhiteundersideofthe
cover.
3.Checkforanystainingontheinteriorfoam.
4.Replaceanystaineditemsanddisposeofasperlocal
authorityprocedure.
5.Removemainspowerandcheckaudiblesounder
operates.
6.Visuallyinspectmainscordforsignsofdamageorwear.
7.2Cleaningandcare
IMPORTANT!
Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedtoclean.
Forfurtherinformationondecontamination
inhealthcareenvironments,pleaserefer
toTheNationalInstituteforClinical
Excellence’guidelinesonInfectionControl
www.nice.org.uk/CG139andyourlocalinfection
controlpolicy.
CleaningHandle
1.TheexterioroftheHandlecanbeperiodicallywiped
usingaclothdampenedwithdisinfectant.
Cleaningcontrolunit
WARNING!
Electricalshockhazard!
Ensurethecontrolunitisdisconnectedfrom
themainselectricitysupplybeforecleaning.
Donotspraydisinfectantdirectlyontothe
controlunit,orimmersethecontrolunitinany
typeofliquid.Thiscouldresultinasevere
electricalhazardasthisequipmenthasno
protectionagainstingressofwater.
Thisequipmentisnotsuitableforuseinthe
presenceofaammableanestheticmixture
withairorwithoxygenornitrousoxide.
161627426-A