Invacare®Softform®PremierActive2S
9Troubleshooting
9.1Identifyingandrepairingdefects
Thereareaudioandvisualalarmspresentonthecontrolunit.
Problem/Alarm
Cause
Solution
MattressCPRhosedisconnected.ConnectCPRhoseconnector,lockitinplace. Mattressnotinating
(notalternating
properly).
Haspowerandfuseisgood,controlunitdoes
notcomeon.
SendcontrolunitbacktoInvacareforrepair.
Majorleakinaircell.Replaceleakingaircell.
CPRhoseortubeconnectorskinkedorsplit.UnkinkorreplacesplitCPRhoseortube
connectors.
Notalternating,rotorfailure.SendcontrolunitbacktoInvacareforrepair.
Noair(controlunitfailure).
SendcontrolunitbacktoInvacareforrepair.
Nopower.
Controlunitoff.Checkpowersource,turnuniton.
Powercorddisconnected.Connectpowercordandensurethepowersource
ison.
Nopowerinthepowersource.Checkthepowersourcehaspowerandturniton.
Poweroutage.
Waittillthepowersourcehaspower.
Fuseblown.Changefuseonpowerinletconnectorwithspare
fuseoridenticalreplacementonly(consulta
trainedengineerifyouareunsurehowtochange
afuse).