EasyManua.ls Logo

ION Panther - Calibration Management

ION Panther
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Panther Instrument User Manual V1.0
Pioneering Gas Sensing Technology. Page 15 of 24 ionscience.com/usa
Calibration Management
To access the Calibration Management on your Panther instrument, click the MENU/SELECT button while on the
main screen to access the Main menu.
Once you are in the Main menu, use the DOWN button and click on ‘Calibration Management by pressing the
‘MENU/SELECT’ button.
Within Calibration Management, you can access the following:
Pump calibration: This option will run the pump calibration. The temperature and
pressure will be measured and used to set the pump to the correct flow rate.
New calibration: This option will firstly provide you with two options,
concentration calibration (ppm) and leak calibration. Please find ‘Concentration
calibration (ppm)’ and ‘Leak calibration’ process below:
Concentration calibration (ppm): Select your chosen gas. To select your chosen
gas, select the first letter of the gas you are looking to detect. Then press ‘DOWN’
to the next option and search for the calibration gas. Once you have found the
calibration, press the MENU/SELECT button.
After you have confirmed the calibration gas, you will need to type in the gas
concentration. Press ‘save as’ and name the calibration. Then press ‘DOWN’ and
select ‘go’ to start calibration.
The instrument will then start the Pump calibration.