EasyManua.ls Logo

ISA T1000 PLUS - Intervention Curve; Ct Saturation Knee Test; Introduction

ISA T1000 PLUS
145 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DOC. MIE91093 Rev. 1.34 Page 136 of 145
restored. Confirm save results pressing the multi-function
knob, and proceed.
1.18.4. Intervention curve
Now we can measure trip timings, following the I-t curve with
as many points as desired.
Now, press ON and pre-adjust the first test current: if the CB
trips, don’t save test result; go OFF. Select ON+TIME: as the
CB trips, test goes OFF; pressing the multi-function knob,
tripping values can be saved. Confirm save results pressing
the multi-function knob, and proceed with other test currents,
until all points to be tested are measured.
1.19 CT SATURATION KNEE TEST
1.19.1. Introduction
The saturation knee test of a CT is performed applying a
voltage to the CT secondary side, and slowly increasing it
while measuring the current sunk by the CT. When the
saturation approaches, the current increases sharply: this is
the saturation knee.
More precisely, the IEC standard C.57.13, paragraph 6.10.2
defines the knee point as the value of the voltage for which a
voltage increase of 10% causes a current increase of 50%”.
T1000 allows to measure and save the applied voltage and the
absorbed current. Repeating the measurements at various
voltage values it is possible to record the I vs. V curve; then,
applying the definition, the knee point can be located
graphically.
NOTE: the test set T3000 performs the measurement in an
automated way.
The voltage to be applied to perform the test is up to 250 V
(500 V for the T1000E model). Before performing the test, you
can compute the nominal knee as follows.

Table of Contents

Other manuals for ISA T1000 PLUS

Related product manuals