GLS Sampler
2-4
The GLS is shipped from the factory with the
following stored program settings:
•Time Paced
15 Minute Pacing Interval
9400 ml Bottle Volume - for 2.5 gallon (10
liter) bottles
Take 96 Samples - covers a 24-hour time
period
80 ml Sample Volume
No Delay to First Sample
3
8-inch by 25-foot suction line
2.4 Standard Programming
Standard Programming allows you to step
through the current GLS program settings and
make any necessary modifications.
To access the Standard Programming, select
Program from the Standby screen and press
Enter.
The seven-step Standard Programming goes
through the following settings:
Pacing - Choose from time or flow paced
sampling. Flow paced sampling requires
an external flow metering instrument.
Interval - Set the pacing interval in
minutes or flow pulses.
Bottle Volume - Enter the capacity of the
bottle installed in the GLS base section.