GLS Sampler
2-12
the GLS advances to the “MAXIMUM RUN
TIME” display.
Some sampling protocols require a known
composite sampling interval, that is, the run
time of a sampling program. For time-paced
sampling programs, the run time is the number
of samples multiplied by the pacing interval.
But for flow-paced sampling, the run time may
vary widely because of the dependency on flow
rates.
If your sampling protocol requires a
known run time, enter the number of
hours for the MAXIMUM RUN TIME. If
the GLS has not yet completed the
sampling program, it will end the
sampling program when it reaches the
the MAXIMUM RUN TIME.
If your sampling protocol does not have
this requirement, enter “0” (zero). The
GLS will run until it completes the total
number of samples.
MAXIMUM RUN TIME: