Balances 19
3.4 IB Series
Follow the steps below to set up and calibrate the Ishida IB Series scale.
CALIBRATION
NOTES:
• The Ishida IB Series precision balance automatically calibrates itself.
• Press the
P key to cancel calibration procedures and return to the weigh display.
• If
[o-Err] appears on the display, the weight exceeds the scale capacity.
• Standard calibration weight is
1/2 capacity. When possible, calibrate using full capacity.
• If
[1-Err] appears on the display, the calibration weight is less than 50% of the standard.
• If
[2-Err] appears on the display, an object other than the calibration weight is on the pan or the display
tolerance (1%) is exceeded.
Step Action Display
1Turn on scale. 0000 0000 0000
2Press and hold the F key until the [CAL] message is displayed. CAL
3Press and hold the T key, press the F key, then release both keys. on 0
4 After zeroing procedure has finished, the following message will be displayed. Set
standard weight in weigh pan. The scale automatically calibrates itself.
on F.S.
5 Scale automatically returns to original measurement mode. 50000