STATus Subsystem
Copyright© Itech Electronic Co., Ltd. 15
Chapter 3 STATus Subsystem
STATus:QUEStionable[:EVENt]?
Queries the event register for the Questionable Status group. This is a
read-only register, which stores (latches) all events that are passed by the
Operation NTR and/or PTR filter. Reading the Questionable Status Event
register clears it.
Syntax
STATus:QUEStionable[:EVENt]?
Arguments
None
Returns
<NR1>
The value returned is the binary-weighted sum of all bits set in the register.
Related Command
STATus:QUEStionable:ENABle
The following table describes the Questionable Status register bit assignments.
Bit
7
6
5
4
3
2
1
0
Name
OT
FAN err
OPP
OCPpeak
OCPrms
UVPrms
OVPpeak
OVPrms
Decimal
Value
128
64
32
16
8
4
2
1
STATus:QUEStionable:CONDition?