Power Supply Function
automatically be OFF and VFD status indicator Prot lights up and the screen
prompts OTP.
Possible Cause
To prevent damaging heat build-up and ensure specified performance, make
sure there is adequate ventilation and air flow around the instrument to ensure
proper cooling. Do not cover the ventilation holes on the rear panel, sides, or
bottom of the instrument. Even with proper ventilation, the instrument can over-
heat in the following situations.
If the ambient temperature is too high.
If you use the instrument to test for long periods.
How to Set
The OTP limit does not need to be set, and the internal device of the instrument
automatically detects and determines whether to enter the OTP state.
If an over-temperature condition occurs, power off the instrument and allow it to
cool for at least 30 minutes. After the internal temperature of the instrument has
cooled down, power it on again.
When you return power to the instrument, verify that the cooling
fan is running. If not, please contact ITECH Technical Support.
Leaving the instrument powered on with an inoperative cooling
fan may result in damage to the instrument.
4.5.7 Sense Reverse Protection
The instrument defaults to provide sense reverse protection. The premise is that
the Sense switch is turned on. When the output state is ON and the difference
between output terminal voltage and sense remote voltage exceeds the speci-
fied voltage, sense reverse protection will be enabled after 500ms. The instru-
ment output will be immediately switched to Off and the display screen will
display SENSE ERR if the sense terminals are reversed.
When the instrument is in Sense Reverse Protection state, you should check
whether the polarities are connected reversely or not firstly. If yes, you can re-
open the output after the polarities connect correctly.
Copyright © Itech Electronic Co., Ltd.
69