Remote Control 
  Copyright © Itech Electronic Co., Ltd.  10 
with the controller (eg PC). Do not use blank Modem cable. Table 2-2 
shows the plug pins. 
If your computer is using a RS-232 interface with DB-25 connector, you 
need an adapter cable with a DB-25 connector at one end and the other 
side is a DB-9(not blank modem cable) 
 
Pin  introduction  of  RS-232 
connector   
 
 
RS-232 Troubleshooting: 
If there is RS-232 connection problem, check the following: 
1.  Computer  and  power  supply  must  configure  the  same  baud  rate, 
parity,  data  bits  and  flow  control  options.  Note  that  the  power 
configuration as a start bit and a stop bit (these values are fixed). 
2.  As described before in RS-232 connector, you must use the correct 
interface cable or adapter. Note that even if the cable has the right 
plug, the internal wiring may be wrong. 
3.  Interface cable must be connected to the correct serial port on the 
computer (COM1, COM2, etc.). 
Communication Settings 
Before communication, you should first make the following parameters 
on the power supply and PC matches. 
Baud  Rate:  9600  (4800,9600,19200,38400,57600,115200).  You  can 
enter the system menu from the front panel, and then set the baud rate. 
  Data bits: 8 
  Stop Bits: 1 
  calibration (none, even, odd) 
  EVEN          8 data bits, have even parity   
  ODD            8 data bits have odd parity 
  NONE          8 data bits, no parity   
  MODE:  (SIG,MUX) 
  SIG    a single instrument is connected to the communication   
  MUX    multiple sets of instrument communication with PC at the same time, 
in this situation, instrument address must be set to distinguish among all 
the units(address    0  ~30, default set is 0)