Compound Control Command 
  Copyright © Itech Electronic Co., Ltd.  26 
Chapter8 Compound  Control  Command 
APPLy  {<voltage  value>|DEF|MIN|MAX}  [{<current 
value>|DEF|MIN|MAX}] 
This order has combined two kinds of orders: VOLTage and CURRent. When 
sending  this  order  to  unit,  power  supply  will  output  voltage  and  current 
according to the current setting of this order. Of course the precondition is the 
setting value is within the set range. If not, then a execution error will occur. You 
can  also use DEF、MIN  or  MAX  as the setting  parameters.DEF represents 
factory setting. MIN will enable the voltage and current to be 0. MAX will enable 
the output to be the highest value in the range. 
Order syntax 
[SOURce:]APPLy <NRf> 
Parameters 
MIN to MAX 
Unit 
V, A 
Query syntax 
APPLy? 
Return parameters 
<NR2>