Chapter6 Current  Control  Commands 
CURRent{<current value>|MIN|MAX|UP|DOWN|DEF} 
This order is used to set the output current. No matter what the current range 
you select currently, the output current is subject to the newest setting value. 
Besides, MIN and MAX can be used as parameters. MIN represents 0A, MAX 
represents the max value of the current range. You may send CURR? MIN or 
CURR? MAX  orders to quire the minimum  and maximum current  within the 
range. DEF represents default value and it also can be used to set value. 
We can also send CURR UP and CURR DOWN orders to increase or decrease 
current, the stepping value should be set with order “CURR:STEP”. 
Oder syntax 
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf> 
Parameters 
MIN to MAX 
Unit 
A 
*RST value 
MAX 
Query syntax 
CURRent?[MINimum|MAXimum] 
Return parameters 
<NR2> 
Example 
CURR:STEP 0.01            //set the stepping value to be 0.01A 
CURR UP                          //enable the output current increase value one time 
CURR:STEP 0.02            //set the stepping value to be 0.02A 
CURR DOWN                  //enable the output current decrease value one time 
CURRent:STEP 
This order is used to service for CURR UP and CURR DOWN two orders. It can 
set  the  stepping  value  of  current.  Besides,  CURR:STEP?  DEF  order  can 
inquire the stepping resolution of current of this unit. For example, the stepping 
setting value is 0.01, it represents 10mA. 
Order syntax 
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement] <NRf>