Calibration Commands
Copyright © Itech Electronic Co., Ltd. 32
Chapter11 Calibration Commands
CALibration:SECure:[STATe]
Set protection mode enable or disable when calibrating the power supply.
Order syntax
CALibration:SECure:[STATe] {<ON|OFF>[<password>]}
Parameters
0|1|ON|OFF 'XXXX
Example
CAL:SEC 1 'XXXX’; //XXXX is the instrument model
CAL:SEC OFF
Query syntax
CALibration:SECure:STATe?
CALibration:INITital
This order is used to reset the calibration parameters to default setting.
Order syntax
CALibration:INITital
Parameters
None
CALibration:SAVe
This order is used to save the calibration parameters in a nonvolatile memory.
Order syntax
CALibration:SAVe
Parameters
None
CALibration:VOLTage:LEVel
This order can set voltage calibration point. P1, P2, P3 and P4 must be
calibrated orderly.
Order syntax
CALibration:VOLTage:LEVel <point>