SCPI Status Register 
  Copyright © Itech Electronic Co., Ltd.  13 
can’t  changed  by  re-setting order (* RST)  or equipment  restart,  but if 
query the data in event register or send clear order *CLS(clear status),   
the event register will be automatically cleared. The main content which 
the standard event register records is: whether power output is turned on, 
order syntax errors, order execution errors, self-test or calibration errors, 
query errors and so on. 
Operation Complete. All commands prior to and 
including an *OPC command have been 
executed. 
Query Error. The power supply tried to read the 
output    buffer but it was empty. Or new 
command line was received before a previous 
query had been read. Or both the input and 
output buffers are full. 
Device Error. A self-test or calibration error 
occurred (see error numbers 601 through 750 in 
chapter 5). 
Execution Error. An execution error occurred (see 
error numbers -211 through -224 in chapter 5). 
Command Error. A command syntax error 
occurred (see error numbers -101 through -178 in 
chapter 5). 
Power On. Power has been turned off and on 
since the   
last time the event register was read or cleared 
 
Query status register provide some information of the power, such as 
over voltage, over temperature, over current. You can also monitor the 
change of the constant current and voltage status via the register, for 
example, data bit 0 is the constant current mode of the power supply, 
data bit 1 is constant voltage mode of the power supply and so on. 
The power supply is/was in the constant 
current mode. 
The power supply is/was in the constant 
voltage mode. 
The fan has a fault condition. 
The  overvoltage  protection  circuit  has 
tripped. 
The  over  current  protection  circuit  has 
tripped. 
 
Status byte register records the information of other registers. The query data is 
temporally  stored in the  output buffer of the  power  supply,  and feedback to