System Commands
Copyright ©ITECH Electronic Co., Ltd. 17
SYSTem:ERRor?
This order is used to query the error information of the power supply. When the
error LED lights on the front panel, it tells us that there is one or more hardware
or order syntax error happened. There are at most 20 groups of error message.
Send the order once will read one error message from the error queue.
Error messages follow the FIFO (first-in-first-out) principle. The first error
will be returned first. when you read all the error tips information in the error
queue. The ERROR led will turn off. When error occur the buzzer of the
power supply will beep once.
If more than 20 errors occurred, the last one stored in the queue will be
replaced by “-350”,
Meaning "too many mistakes.". If the error
messages in the error message queue are not read out, the error
following will not be saved to the queue. If there is no error message,
when read error information, it will return “+0”, meaning “no error”.
If turn off the power supply or send *CLS (clear status) order, the error
messages in the error queue will be cleared. *RST order will not clear the
error messages in the error queue.
Order syntax:
SYSTem:ERRor?
Parameter:
None
SYSTem:REMote
This order is used to set the power supply to remote control mode through the
RS232 interface. Except for the Local key on the front panel, other keys are
locked and can’t be used. Send or receive orders without first sending the order
to configure the remote control may cause unpredictable results.
Order syntax:
SYST:REM
Parameter:
None
Quest syntax:
None
SYSTem:LOCal
This order is used to set the power supply to panel control mode through the
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com