Installation 
  Copyright ©ITECH Electronic Co., Ltd. 17 
Chapter3 Installation 
3.1 Connectiong the Power Cord 
Connect power cord of standard accessories and ensure that the power supply 
is under normal power supply. 
Before connecting the power cord 
To prevent electric shock and damage to the instrument, observe the following 
precautions. 
 
z  Before connecting power cord, be sure to confirm that the power voltage 
matches with the rated input voltage of the instrument. 
z  Before connecting power cord, be sure to switch off the instrument. Verify 
that there is no dangerous voltage on the connection terminals. 
z  To avoid fire or electric shock, Make sure to use the power cord supplied by 
ITECH. 
z  Be sure to connect the power cord to the AC distribution box with protective 
grounding. Do not use terminal board without protective grounding. 
z  Do not use an extended power cord without protective grounding, otherwise 
the protection function will fail. 
z  Ensure that the power cord connection terminals are either insulated or 
covered by the supplied protective cover so that no accidental contact with 
lethal voltage can occur. 
 
 
 
Safety agency requirements dictate that there must be a way to physically 
disconnect the AC mains cable from the unit. A disconnect device, either a 
switch or circuit breaker must be provided in the final installation. The 
disconnect device must be close to the equipment, be easily accessible, and 
be marked as the disconnect device for this equipment. 
 
Connecting the power cord 
The standard power cord specifications for this series of 3U instruments are 
divided into the following types according to different regions: