Input Settings
Copyright © ITECH Electronic Co., Ltd. 26
[SOURce:]CURRent:SLEW[:BOTH] <NRf+>
This command sets the slew rate for all programmed changes in the input
current level of the electronic load. This command programs both positive and
negative going slew rates.
Command Syntax
[SOURce:]CURRent:SLEW[:BOTH] <NRf+>
Parameters
MIN ~MAX | MAXimum | MINimum
Unit
A /uS
*RST Value
MAXimum
Examples
CURR:SLEW 3
Related Commands
[SOURce:]CURRent:SLEW?
Returned Parameters
<NR2>
[SOURce:]CURRent:SLEW[:BOTH]?
This command is used to query the slew rate for all programmed changes in the
input current level of the electronic load.
Command Syntax
[SOURce:]CURRent:SLEW[:BOTH]?
Unit
A /uS
*RST Value
MAXimum
[SOURce:]CURRent:SLEW:RISE <NRf+>
This command sets the slew rate of the current for positive going transitions.
Command Syntax
[SOURce:]CURRent:SLEW:RISE <NRf+>