Status Commands
Copyright © Itech Electronic Co., Ltd. 51
Returned value:
<NR1>
Related command:
STATus:OPERation:ENABle
The bit configuration of the Operation status registers is as follows:
Bit 0 Calibrating
Bit 5 Trigger
STATus:OPERation:CONDition?
This query returns the value of the Operation Condition register. That is a
read-only register, which holds the live (unlatched) operational status of the
electronic load.
Query Syntax:
STATus:OPERation:CONDition?
Parameter:
None
Returned value:
<NR1>
STATus:OPERation:ENABle
This command set the value of the Operational Enable register. This register is
a mask for enabling specific bits from the Operation Event register to set the
operation summary bit (OPER) of the Status Byte register.
Command syntax:
STATus:OPERation:ENABle <NR1>
Parameter:
0~65535
Example:
STATus:OPERation:ENABle 128
Query Syntax:
STATus:OPERation:ENABle?
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com