9. Product Specifications
9.1 Delivery form
Programat CS
Power cord
Vacuum hose
Cooling Tray “S”
Programat Firing tongs
Programat Firing Tray Kit
Automatic Temperature Calibration Set ATK2 (Test Set)
USB data cable
PrograBase 2 Software
Service passport equipment
Operating Instructions
Recommended accessories
Programat Accessories Set
Automatic Temperature Checking Set 2 (ATK 2)
Vacuum Pump VP3 easy
9.2 Technical data
Power supply 100 V / 50–60 Hz
110–120 V / 50–60 Hz
200–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Tolerated voltage fluctuations +/- 10%
Max. power consumption 11 A at 100 V
12 A at 110–120 V
8.5 A at 200–240 V
Acceptable data for vacuum pump of other manufacturers
Max. output: 250 W / max. leakage current 0.75 mA
Final vacuum: < 50 mbar
Use only tested pumps
Electrical fuses: 100 V / 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
200–240 V:
250 V / T 8 A ( heating circuit)
250 V / T 3.15 A ( vacuum pump)
Dimensions of electrical fuses 110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Dimensions of the closed furnace
Depth: 430 mm / width: 305 mm / 410 mm (with Cooling tray)
Height: 320 mm
Usable size of the firing chamber Diameter 80 mm
Height 48 mm
Max. firing temperature 1200 °C (2192 °F)
Weight Furnace base: 12.0 kg
Furnace head: 4.5 kg
Safety information
The CS complies with the following guidelines:
IEC 1010-1/EN 61010, Part 1
– UL and cUL standards
Radio protection / electromagnetic compatibility EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 ° F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m above sea
level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range -20 to +65 °C (-4 °F to +149 °F)
Acceptable humidity range Max. 80 % relative humidity
Acceptable ambient pressure 500 mbar to 1060 mbar
Use only original packaging of the Programat CS together with the
respective foam material for shipping purposes.