EasyManua.ls Logo

Jadever SKY - Panel Illustration; Single Point Calibration of Weight

Jadever SKY
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electronic Scale
3
Panel Illustration
1
Level
6
Stable indicator
2
Warming up indicator
7
Tare key
3
Sampling/Backlight key
8
Zero key
4
Function selection key
9
Weight display window
5
Unit conversion key
Single Point Calibration of
Weight
Step 1: Power off while holding down MODEkey, and then press ON/OFFto switch
on the scale. The screen will show the following:
Count
Step 2: Press the MODEkey, the screen will show the following:
CAL
Step 3: Press the UNITkey to start zero point calibration. Make sure there is nothing on
the weighing pan. The screen will show the following:
ZEro
Please wait for the screen to show: