EasyManua.ls Logo

Jadever SKY - Linearity Calibration of Weight

Jadever SKY
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electronic Scale
4
CAL-S
The calibration can be done at the point of 1/3 load, 2/3 load or full load. Choose the one
you need and place the corresponding weight onto the pan.
If place 1/3 of load onto the pan, the
screen will show:
on1
If place 2/3 of load onto the pan, the
screen will show:
on2
If place full load onto the pan, the
screen will show:
on3
Step 4: Place weights of equivalent weight onto the pan, wait for a beep sound and the
screen will show:
PASS
Step 5: Press MODEand then press SMPL, the screen will show:
busy
Step 6: After countdown, the single point calibration is completed and the scale returns to
the normal weighing mode.
Linearity Calibration of Weight
Step 1: Power off while holding down ZEROand TAREkeys, and then press ON/OFF
to switch on the scale. The screen will show the following:
L-CAL
Step 2: Clear the weighing pan, then press TAREand the screen will show the following:
(zero point calibration)
on0
Step 3: Wait for a beep sound and the screen will show the following:
on1
Step 4: Place 1/3 of load on the weighing pan in ten seconds ,Waiting for a beep sound and
the screen will show the following:
on2
Step 5: Place 2/3 of load on the weighing pan in ten seconds ,Waiting for a beep sound and
the screen will show the following:
on3
Step 6: Place full load on the weighing pan in ten seconds, Waiting for a beep sound and the
screen will show the following:
PASS
Step 7: Clear the weighing pan, then press TAREand the screen will show the following:
busy

Related product manuals