Electronic Scale
4
CAL-S
The calibration can be done at the point of 1/3 load, 2/3 load or full load. Choose the one
you need and place the corresponding weight onto the pan.
If place 1/3 of load onto the pan, the
screen will show:
on1
If place 2/3 of load onto the pan, the
screen will show:
on2
If place full load onto the pan, the
screen will show:
on3
Step 4: Place weights of equivalent weight onto the pan, wait for a beep sound and the
screen will show:
PASS
Step 5: Press MODEand then press SMPL, the screen will show:
busy
Step 6: After countdown, the single point calibration is completed and the scale returns to
the normal weighing mode.
Linearity Calibration of Weight
Step 1: Power off while holding down ZEROand TAREkeys, and then press ON/OFF
to switch on the scale. The screen will show the following:
L-CAL
Step 2: Clear the weighing pan, then press TAREand the screen will show the following:
(zero point calibration)
on0
Step 3: Wait for a beep sound and the screen will show the following:
on1
Step 4: Place 1/3 of load on the weighing pan in ten seconds ,Waiting for a beep sound and
the screen will show the following:
on2
Step 5: Place 2/3 of load on the weighing pan in ten seconds ,Waiting for a beep sound and
the screen will show the following:
on3
Step 6: Place full load on the weighing pan in ten seconds, Waiting for a beep sound and the
screen will show the following:
PASS
Step 7: Clear the weighing pan, then press TAREand the screen will show the following:
busy