GUIDELINES AND MANUFACTURER' S STATEMENT-ELECTROMAGNETIC IMMUNITY
Anti-interference
measurement
IEC60601
Test Level
RFconduction
RF radiation
(charger)
RF radiation
(wheelchair)
3 Vrms
3 Vrms
150 kHz to
80 MHz
3 V/m
3 V/m
80 MHz to
1.0 GHz
20 V/m
20 V/m
26 MHz to
2.5 GHz
Recommended isolation distance
80 MHz to 800 MHz
800 MHz to 1.0GHz
d=
P0.2
d=
P0.4
d=
P1.2
d=
P1.2
d=
P2.3
26 MHz to 800 MHz
800 MHz to 2.5 GHz
Coincidence
level
Electromagnetic
environment-Guidelines
ISO7176
IEC60601
ISO7176
IEC60601
ISO7176
IEC60601
D26
Power Wheelchair
is expected to be used in the following specified electromag
-
netic environment, and the purchasers or users of NPL001
NPL002
NPL003
electric wheelchairs shall ensure that it is used in this electromagnetic environment
Note 1: at the frequency of 80MHz and 800MHz, the formula of higher frequency band
is adopted.
Note 2: these guidelines may not be suitable for all cases where electromagnetic prop
agation is affected by absorption and reflection of buildings, objects and human bodies.
Portable and mobile RF communi
-
cation equipment shall not be used
closer to any part of the electric
wheelchair, including cables, than
the recommended isolation dis
-
tance. The distance shall be calcu
lated by the formula corresponding
to the transmitter frequency.
Where P is the maximum output
rated power of the transmitter pro
-
vided by the transmitter manufactur
-
er, in watts (W), and d is the recom
-
mended isolation distance in meters
(m).
B
The field strength of the fixed RF
-
transmitter is determined by survey
-
ing the electromagnetic field A. in
each frequency range, B should be
lower than the coincidence level.
Interference may occur near equip
-
ment marked with the following
symbols.