JDSU ORL-55 57
8 R
EMOTE CONTROL
:SOUR
:WAV
:AUTO
Sets Auto- modulation ON/OFF.
Parameter type: <BOOLEAN>
•0: Auto- OFF
•1: Auto- ON
Default setting: OFF (0)
:SOUR
:WAV
:AUTO?
Returns Auto- modulation status.
•0: Auto- OFF
•1: Auto- ON
:SOUR
:MOD
Sets modulation for the selected
LASER source.
Parameter type: <NR1>
Values allowed:
0: CW (continuous wave)
270:mod. freq. 270 Hz
1000:mod. freq. 1000 Hz
2000:mod. freq. 2000 Hz
Note: Setting fails, if more than one
LASER source is selected.
:SOUR
:MOD?
Returns current modulation.
0: CW (continuous wave)
270:mod. freq. 270 Hz
1000:mod. freq. 1000 Hz
2000:mod. freq. 2000 Hz
AUTO:Auto modulation is ON.
:SOUR
:POW
Sets the power level for the specified
LASER source.
Parameter type: <NR1,NR1>
e.g. :SOUR:POW 1,-522
sets the power level for LASER1 to -
5.22 dBm
Command
string
Parameter type / Response type /
Unit / Info