58 ORL-55 JDSU
8 R
EMOTE CONTROL
Power Meter commands
:SOUR
:POW?
Returns the power level [dBm x 100]
for the specified LASER source.
Parameter type: <NR1>
e.g. :SOUR:POW? 1
:SOUR
:STAT
Sets state of the LASER ON/OFF.
Parameter type: <BOOLEAN>
0: LASER OFF
1: LASER ON
:SOUR
:STAT?
Returns the state of the LASER.
0: LASER OFF
1: LASER ON
Command
string
Parameter type / Response type /
Unit / Info
:POW
:CAL
:WAV
:MAX?
Returns the maximum wavelength
that can be set.
e.g. 1650
:POW
:CAL
:WAV
:MIN?
Returns the minimum wavelength
that can be set.
e.g. 780
:POW
:CAL
:WAV
:TAB
Sets up to 30 wavelengths in the
-Table.
Parameter type: <NR1,NR1,...>
e.g. :POW:CAL:WAV:TAB 820,850,
980,1280,1300,1310,1480,1510,
1550,1625
Command
string
Parameter type / Response type /
Unit / Info