ERROR CODES
ERROR
CODE
DESCRIPTION
PROBABLE CAUSE
SHORTED
THERMOCOUPLE
ORDEFECTIVE
HEATERPLATES
Occursifthethermocouplewiresareshortedortheheaterplatesare
defectivecausingInfinitytoachieveaHEATRATEoflessthan6%ofthe
maximumattainableHEATRATEwithfullpowerappliedtotheheater
platesfor5minutesThiserrorwillalsooccurifaprogramisrunning
andthedooriskeptopenformorethan5minutesorthetriacorthe
MainPCboarddevelopedaproblemduringaburnoutprocess
Er6
THERMAL
RUNAWAY
Er7
INVALIDENTRY
ERRORWHILEA
PROGRAMIS
RUNNING
Flashing
99:99
when
NIGHT
TIME
(DELAY
START)is
pressed
duringa
program
OccurswhentheHEATRATEinSTAGE2,3or4issetto“COOL”but
thecorrespondingTEMPisenteredtoheattoahighertemperature
thanthepreviousstage
Occurswhenthetemperaturehasexceeded2015°F(1102°C)for1
minuteifthehighesttemperatureprogrammedwaslessthan2000°F
(1093°C)Thisalsooccurswhenthetemperaturehasexceeded
2050°F(1121°C)for1minuteifthehighesttemperatureprogrammed
was2000°F(1093°C)orhigher
18