EasyManua.ls Logo

Kaixin MSU2 - Slope;Heart Rate;Cycle Measurement

Kaixin MSU2
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments (Vet) Users Manual V2.14
-
17
-
Circumference/area measurement with trace method
1. In B, B/B mode, freeze desired image, the cursor is located in the “Measureposition of display
area;
2. Press key, the measurement methods are showed in the lower left of the screen, press direction
keys to chooseTrace”, press key again, the cursor will show +”;
3. Press direction keys to move the “+ mark to desired position, press key to set the +mark
position as the starting point of the measurement;
4. Press direction keys to move the“+mark to the end point of the measurement. At the same time, a
locus appears in the direction of operation between the two measurement marks. The measured
circumference value is displayed automatically at the built-in mark C: 00000mm on the right
part of the screen. Press key to display at the built-in mark A: 00000mm
2
the value of the
measured area formed by measurement line enclosure;
5. Press key, the measurement methods are showed in the lower left of the screen, continue to
chooseTrace” to measure;
6. Repeat steps 3, 4 to complete the multi-group data measurement;
7. Continuously press key twice to quit the measurement status;
8. Press direction keys to choose “Clear” and then press key to clear all marks and data;
9. Press key to unfreeze, clear all marks and data and quit the measurement status.
7.3 Slope/Heart rate/Cycle Measurement
The method to measure slope/hear rate/cycle is identical with distance measurement.
Note: In B/M mode, if both starting point and end point of the measurement mark fall into the
B-mode area, the value of the+:refers to distance; if starting point and end point of the
measurement mark fall into the M-mode area, the value of the+:refers to depth; if the
starting point and end point are in separate areas, the+:will display----”sign or
invalid value.
: denotes depth measured in mm (millimeter)
EF: denotes slope coefficient measured in mm/s (millimeter per second)
HR: denotes heart rate measured in times/minute (times per minute)
T: denotes cycle measured in ms (millisecond)
Attention: The accuracy of software measurement: distance measurement 1mm;
area measurement 1mm
2
; volume measurement 1cm
3
; heart rate measurement
1bpm; time measurement ≤ 1ms. Due to differences in images obtained by each user in
different times, the actual object for the accuracy of the measurement may be greater
than the above-mentioned values.