7. WHEN YOU FINISH USING THE ANALYSER
Always try to switch the analyser off in fresh air .
The analyser counts down from 30 before switch off with the pump running to self
clean its sensors. If the measured CO level is more than 30 ppm do not switch off the
analyser. It will need to sample fresh air until the CO level falls below 30 ppm. This
extends the sensor's life and helps to eliminate any unwanted offset on next start-up.
8. PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an authorized repair
centre.
Fault symptom Causes / Solutions
Batteries not holding charge. Batteries exhausted.
AC charger not giving correct output.
Fuse blown in charger plug.
Analyser does not respond to
gas
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with
contaminants.
Water Trap Bung not fitting tightly.
Sensor fault
TROUBLE SHOOTING
I have set the analyser to AUTOLOG but it has stopped logging:
Memory might be full
Menu
Memory
View ….. displayed number is last log number. 255 is full
PUMP is OFF
Analyser is PRINTING