Page 16
English 18259
Display is totally clear or totally dark
Set the rotary dial to MENU switch on and leave unit for five minutes.
Press UP three times (listen for three beeps) then press ENTER
Press and hold down UP until the display becomes visible again (this can take
up to 20 seconds). Use UP and DOWN to adjust the display's contrast.
Press ENTER to exit.
The analyser continually beeps and the display has frozen.
You have exited a menu function without completing the required sequence.
Rotate the dial back to MENU
Press ENTER until MENU appears on the top line
- (NB You may be changing settings when you do this)
Rotate the dial to the required display position.
The unit will not turn off.
Do not hold down the on/off button.
Try connecting the charger.
9. ANNUAL RECALIBRATION AND SERVICE
Although sensor life is typically more than two years the analyser should be re-
calibrated and serviced annually to stop any long-term sensor or electronics drift or
accidental damage.
Local regulations may require more frequent re-calibration.
SERVICE MODE
This mode can be selected from MENU and gives password protected access to
certain functions WARNING: entering inappropriate codes may corrupt the
calibration of this instrument.
The User level pass code to view software issue is 1111