4.5 ROOM CO TESTING
Select “Room CO” to measure and record CO readings for up to 30 minutes.
→
→
→
→
Use / to select the test type from
the following:
TEST TYPES
GENERAL: 15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
SWEEP TEST: 2 minute test with max reading
stored at end
LIMIT = 10ppm
ALARM = 30 ppm
MIGRATION TEST: 15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE C SEALED APPLIANCE: 15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE B BOILER OPEN FLUE: 15 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE A COOKER: 30 minute test with results
stored every miute
LIMIT = 30ppm
ALARM = 90 ppm
TYPE A WATER HEATER: 5 minute test with results
stored every minute
LIMIT = 10ppm
ALARM = 30 ppm
TYPE A SPACE HEATER: 30 minute test with results
stored every 1 minute
LIMIT = 10ppm
ALARM = 30 ppm
Press / to start Room CO testing.