▶ Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶ Then rinse for 3-5 seconds with KaVo DRYspray.
See also: Instructions for use for the KaVo CLEANspray/KaVo DRYspray.
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are
only available in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxem‐
bourg, Belgium, Netherlands, United Kingdom, Denmark, Sweden, Fin‐
land and Norway.
In other countries,only automated interior cleaning with thermodisinfectors
in accordance with ISO 15883-1.
Preparation methods according to ISO 17664 35