Depending on the device, select a suitable method from the following grav‐
itation methods.
Autoclave with a triple pre-vacuum for least four minutes at 134°C ± 1°C
(273°F ± 1.8°F)
Autoclave using the gravitational method: sterilise for at least 10 min. at
134°C ± 1°C (273°F ± 1.8°F)
Autoclave using the gravitational for at least 60 min. at 121°C ± 1°C (250°F
± 1.8°F)
Follow the manufacturer's instructions for use.
6.8 Storage
Prepared products must be stored protected germ-free from dust in a dry,
dark and cool room.
Preparation methods according to ISO 17664 49