Instructions for use PROPHYflex 3 - 2018 for Sirona connection - 1.006.9920, - 1.006.9927
7 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection
20 / 26
7.2.3 Manual cleaning of the inside
Validated internal cleaning (removal of residual protein) can be accomplished with
KaVo CLEANspray and KaVo DRYspray.
▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the cor-
responding servicing adapter. Press the spray button three times for 2 seconds
each time. Remove the medical device from the spray attachment and let the
cleanser act for one minute.
▶ Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also:
2 KaVo CLEANspray / KaVo DRYspray Instructions for Use.
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only avail-
able in the following countries:
Belgium, Denmark, Germany, Finland, France, United Kingdom, Italy, Luxembourg,
Netherlands, Norway, Austria, Poland, Portugal, Sweden, Switzerland and Spain.
In other countries interior cleaning can only be carried out with washer disinfectors
in accordance with EN ISO 15883-1.
7.2.4 Automated internal cleaning
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.
▶ In order to prevent damage to the medical device due to residual fluid, make sure
that the inside and outside of the device is dry after the end of the cycle. Remove
any residual liquids from the interior and exterior of the medical device using com-
pressed air.
The drying procedure is normally part of the cleaning program of the thermodisin-
fector.
Note
Please comply with the instructions for use of the washer disinfector.
7.3 Disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶ Principally, KaVo recommends carrying out an final disinfection of the unpack-
aged item to guarantee the complete disinfection.
NOTICE
Never disinfect the handpiece with chlorine-containing products.
Malfunction and material damage.
▶ Only disinfect in a washer disinfector or manually.