Table 6. CONFIGURE OHMS-2W menu structure
Menu item Description
SPEED
NORMAL
FAST
MEDIUM
HIACCURACY
SET-SPEED-EXACTLY
SET-BY-RSLN
Measurement speed (integration time)
menu:
Select 1 PLC (power line cycle,
16.67msec for 60Hz, 20msec for
50Hz and 400Hz).
Select 0.01 PLC.
Select 0.1 PLC.
Select 10 PLC.
Set integration time in PLC
(0.01-10).
Default to setting appropriate for
resolution.
FILTER
AUTO
AVERAGING
ADVANCED
AVERAGING-MODE
Digital filter menu:
Default to filter appropriate for inte-
gration time.
Program a simple average filter
(1-100 readings).
Program a simple average filter
(1-100 readings) with a noise tol-
erance window (0-100% of
range).
Select moving average or repeating
average mode.
RESOLUTION
AUTO
3.5d, 4.5d, 5.5d, 6.5d,
7.5d
Display resolution menu:
Default to resolution appropriate for
integration time.
Select a specific resolution.
OFFSETCOMP Enable/disable offset compensation
(20
Ω
-20k
Ω
ranges).
MAXAUTORANGE
1G
Ω
, 200M
Ω
, 20M
Ω
,
2M
Ω
, 200k
Ω
, 20k
Ω
Set upper limit for autoranging of 2-
wire resistance.
Select a specific range.