48
Error and Status Messages
Table 20. Error and status messages
Number Description Event
+900 “Internal System Error” EE
+611
+610
“Questionable Temperature”
“Questionable Calibration”
SE
SE
+515
+514
+513
+512
+511
+510
“Calibration dates lost”
“DC calibration data lost”
“AC calibration data lost”
“Power-on state lost”
“GPIB address lost”
“Reading buffer data lost”
EE
EE
EE
EE
EE
EE
+444 to
+350
Calibration Errors
(see Calibration Manual)
EE
+312
+310
+309
+308
+306
+305
+304
+303
+302
+301
“Buffer Pretriggered”
“Buffer full”
“Buffer half full”
“Buffer Available”
“Reading Available”
“High limit 2 event”
“Low limit 2 event”
“High limit 1 event”
“Low limit 1 event”
“Reading overflow”
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
+174
+173
+172
+171
“Re-entering the idle layer”
“Waiting in arm layer 2”
“Waiting in arm layer 1”
“Waiting in trigger layer”
SE
SE
SE
SE
+161 “Program running” SE
+126
+125
+124
+123
+122
+121
“Device calculating”
“Device measuring”
“Device sweeping”
“Device ranging”
“Device settling”
“Device calibrating”
SE
SE
SE
SE
SE
SE
+101 “Operation Complete” SE
+000 “No Error” SE

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz