55
Status Summary Messages
* STB ?
Serial Poll
OSB
(B7)
RQS
(B6)
MSS
ESB
(B5)
MAV
(B4)
QSB
(B3)
EAV
(B2) (B1) (B0)
OR
* SRE
* SRE ?
Status Byte
Register
Service Request
Enable Register
OSB = Operation Summary Bit
MSS = Master Summary Status
RQS = Request for Service
ESB = Event Summary Bit
MAV = Message Available
QSB = Questionable Summary Bit
EAV = Error Available
MSB = Measurement Summary Bit
& = Logical AND
OR = Logical OR
OSB
(B7) (B6)
ESB
(B5)
MAV
(B4)
QSB
(B3)
EAV
(B2) (B1) (B0)
&
&
&
&
&
MSB
MSB
&
Service
Request
Generation
Read by Serial Poll
Read by *STB ?
Figure 6. Status byte and service request (SRQ)

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz