67
:CURRent:AC
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:COUPling AC|DC
:COUPling?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
Path to configure AC current.
Specify integration rate in sec-
onds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable
(0 or OFF) auto aperture.
Enable and then disable
auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration
rate).
Specify integration rate (num-
ber of line cycles; 0.01 to
10).
Enable (1 or ON) or disable
(0 or OFF) auto NPLC.
Enable and then disable
auto NPLC.
Query auto line cycle inte-
gration (0 or 1).
Query line cycle integration
rate.
Specify input coupling.
Query input coupling.
Path to configure measure-
ment range:
Select range (0 to +2.1).
Query range.
Enable (1 or ON) or disable
(0 or OFF) auto range.