4: Remote operation Model 2290-5 5 kV Power Supply
User's Manual
4-6 2290-5-900-01 Rev. A/December 2013
SMOD?
The SMOD? query returns the VSET setting mode. A returned value of 0 indicates that the voltage
value is controlled by the front panel or bus setting, while the value 1 indicates that the output is
controlled by the rear panel SET/MON voltage control input. Note that the setting mode may only be
changed by setting the rear panel VOLTAGE switch.
TCLR
The TCLR command clears any voltage or current trips.
TMOD(?) <n>
The TMOD command sets the trip reset mode. The value <n> = 0 sets manual trip reset, while the
value <n> = 1 sets the trip reset mode to automatic.
VLIM(?) <n>
The VLIM command sets the value of the voltage limit to <n>, where <n> is in volts. The VLIM?
query returns the present limit setting. As with front panel control, the VLIM value must be greater
than or equal to the VSET value, or an execution error will occur.
Example:
VLIM 500
Set 500 V voltage limit
Default value: 5000 V
VSET(?) <n>
The VSET command sets the voltage output level to <n>. If rear panel control is enabled, an error is
returned. The value <n> is in volts. The VSET? query returns the current VSET value. As with front
panel control, the VSET value must be less than or equal to the VLIM value, or an execution error
will occur.
Example:
VSET 2500
Set output to 2.5 kV
Default value: 0 V