Model 2290
-5 5 kV Power Supply User's Manual Section 4:
Remote operation
2290-5-900-01 Rev. A/December 2013 4-5
VOUT?
The VOUT? query returns the actual output voltage. This value is the same as that shown on the front
panel VOLTS display. The value returned is a floating-point number and includes the positive sign of
the output voltage.
As with the front panel voltage display, this value takes about one second to stabilize after a change
in voltage.
Setting control commands
*RCL <n>
The *RCL (Recall) command recalls stored configuration setting <n>, where <n> ranges from 0 to 9.
Setting 0 recalls the default settings. If the stored setting is corrupted, an error is returned.
Example:
*RCL 0
Recall default setting
*SAV <n>
The *SAV (Save) command stores the present instrument setup as setting <n>, where <n> ranges
from 1 to 9.
Example:
*SAV 3
Save setting #3
ILIM(?) <n>
The ILIM command sets the value of the current limit to <n>, where <n> is a floating-point value in
amperes. The ILIM? query returns the current limit setting.
Example:
ILIM 1E-3
Set 1 mA current limit
Default value: 5.25 mA
ITRP(?) <n>
The ITRP command sets the value of the current trip to <n>, where <n> is a floating-point value in
amperes. The ITRP? query returns the current trip setting.
Example:
ITRP 1E-3
Set 1mA trip current
Default value: 5.25 mA